FLOW-SAFE II EZ CPAP with Nebulization
- Over 50-Percent less oxygen consumption: Uses less oxygen while delivering high Fi02
- Built in manometer: Verifies delivered CPAP pressure
- Pressure relief valve: Automatically adjusts to avoid excess pressure
- Nebulizer built in: Clinicians can administer meds through the integrated EZflow MAX Nebulizer
- Advanced mask design: Lightweight contoured mask and elastic head harness provide a better seal and comfort.
Are there just two sizes of CPAP masks available with the FLOW-SAFE® II EZ CPAP with Nebulization?
Anonymous- On Thursday, 25 March 2021
No, the FLOW-SAFE® II EZ CPAP with Nebulization is also available with a child size mask by special order. For ordering information, please call Customer Care at 800-824-6016, and reference part number MERC_1057320.
Life-Assist- answered On Friday, 26 March 2021
What is the approximate weight of one single packaged FLOW-SAFE II EZ CPAP with Nebulization?
Anonymous- On Monday, 31 January 2022
The approximate weight of one single packaged FLOW-SAFE II EZ CPAP with Nebulization is 0.65 pounds.
Forrest U.- answered On Tuesday, 01 February 2022
Can the Neb system on the FLOW-SAFE II EZ CPAP with Nebulization be purchased separately and used with the other FLOW-SAFE II CPAP or BiPAP systems?
Anonymous- On Friday, 05 January 2024
No, the neb system on the FLOW-SAFE II EZ CPAP with Nebulization cannot be purchased separately.
Jim L.- answered On Friday, 05 January 2024
Is there any literature or explanation on the pressure relief valve on the FLOW-SAFE II EZ CPAP with Nebulization? All I can find is: "Automatically adjusts to avoid excess pressure" but we have a step in our use policy to "verify pressure relief active" and I do not know how to mitigate satisfaction of this without any literature behind what the mechanism is. Thank you!
Anonymous- On Sunday, 24 March 2024
According to the manufacturer, the FLOW-SAFE II EZ CPAP with Nebulization pressure relief is a passive valve that opens when the pressure exceeds 25 cmH2O. The pressure relief valve is mentioned in the IFU. Please call Customer Care at 800-824-6016 if you would like a copy of the IFU.
Jim L.- answered On Tuesday, 26 March 2024
Does the FLOW-SAFE II EZ CPAP with Nebulization nebulizer use the same O2 line? We are looking for a cpap that will only require 1 line or 1 02 bottle
Andrew L.- 4 weeks ago
The nebulizer connection on the FLOW-SAFE II EZ CPAP with Nebulization is integrated. It only requires one O2 tube.
Jim L.- answered 4 weeks ago